Manuals: [https://clonwiki0.jlab.org/wiki/clondocs/man/VME/V1290_REV5.pdf REV5(pdf)], [https://clonwiki0.jlab.org/wiki/clondocs/man/VME/V1290_REV12.pdf REV12(pdf)], [https://clonwiki0.jlab.org/wiki/clondocs/man/VME/addendumV1190.pdf 2eSST addendum]  
Manuals: [https://clonwiki0.jlab.org/wiki/clondocs/man/VME/V1290_REV5.pdf REV5(pdf)], [https://clonwiki0.jlab.org/wiki/clondocs/man/VME/V1290_REV12.pdf REV12(pdf)], [https://clonwiki0.jlab.org/wiki/clondocs/man/VME/addendumV1190.pdf 2eSST addendum]  
== Calibration procedure ==





Revision as of 08:41, 26 July 2018

Manuals: REV5(pdf), REV12(pdf), 2eSST addendum


Calibration procedure

Test results : Calibration VX1290A with DSC2

TDC Outputs from JLAB VME Discriminators DSC2 were Inputs for CAEN VX1290A.

Width of TDC Outputs changed from 4 ns to 40 ns.

Histograms present distributions for four CAEN VX1290A TDCs in each run

Test was done in Counting room VME crate "croctest2"


  • In run # 10491 four VX1290A with Ser. ## 56, 60, 53, 55 were tested, TDCs were in slots 7, 8, 9, 10 correspondingly:

1290.Ser.56.60.53.55_10491.pdf

1290.Ser.56.60.53.55_10491.ps


  • Run # 10493, four VX1290A, Ser. ## 49 34 39 37 in slots ## 7 8 9 10:

1290.Ser.49.34.39.37_10493.pdf

1290.Ser.49.34.39.37_10493.ps


  • Run # 10494, four VX1290A, Ser. ## 21 23 38 52 in slots ## 7 8 9 10:

1290.Ser.21.23.38.52_10494.pdf

1290.Ser.21.23.38.52_10494.ps


  • Run # 10501, four VX1290A, Ser. ## 27 26 32 43 in slots ## 7 8 9 10:

1290.Ser.27.26.32.43_10501.pdf

1290.Ser.27.26.32.43_10501.ps


  • Run # 10505, four VX1290A, Ser. ## 44 50 51 33 in slots ## 7 8 9 10:

1290.Ser.44.50.51.33_10505.pdf

1290.Ser.44.50.51.33_10505.ps



Tests of CAEN TDCs VX1290N with NIM inputs

TDC Outputs from JLAB VME Discriminators DSC2 were converted on ECL-to-NIM translator and were used as Inputs for CAEN VX1290N. CAEN TDCs VX1290N and JLAB VME Discriminators DSC2 were tested in Counting room, in VME crate "croctest2".

Width of TDC Outputs changed from 4 ns to 40 ns.

Histograms suppose to present distributions for four CAEN VX1290N TDCs in each run, but in run 10514 only three VX1290N were tested. In run 10516 - fourth TDC with Ser.28 was not visible in crate, something wrong with setting of base address.


  • In run # 10514 three VX1290N with original firmware v.0.5 and Ser. ## 22, 23, 24 were tested, TDCs were in slots 7, 8, 9 correspondingly:

1290n.Ser.22.23.24.00_10514.pdf

1290n.Ser.22.23.24.00_10514.ps


  • Run # 10516 for four VX1290N with firmware v.0.6, Ser. ## 25 29 31 28 - slots ## 7 8 9 10, Ser.28 - had a problem, not visible in crate:

1290n.Ser.25.29.31.28_10516.pdf

1290n.Ser.25.29.31.28_10516.ps


  • Firmware changed to v.0.6 for Ser. 22, 23 and 24. Address Cards were changed for Ser.28, now VX1290N Ser.28 PBCM0044 is visible in crate.

So, in run # 10520 four VX1290N with firmware v.0.6 and Ser. ## 22, 23, 24, 28 were tested, TDCs were in slots 7, 8, 9, 10 correspondingly:

1290n.Ser.22.23.24.28_10520.pdf