CAEN V1290

From CLONWiki
Jump to navigation Jump to search

Manual (pdf)


Test results : Calibration with DSC2

TDC Outputs from JLAB VME Discriminators DSC2 was an Input for CAEN VX1290A.

Width of TDC Outputs changed from 4 ns to 40 ns.

Histograms present distributions for four CAEN VX1290A TDCs in each run

Test was done in Counting room VME crate "croctest2"


In run # 10491 four VX1290A with Ser. ## 56, 60, 53, 55 were tested, TDCs were in slots 7, 8, 9, 10 correspondingly:

1290.Ser.56.60.53.55_10491.pdf 1290.Ser.56.60.53.55_10491.ps


1290.Ser.56.60.53.55_10491.pdf 1290.Ser.56.60.53.55_10491.ps



  • FADC250 Signal: one channel - 100 events; width because of trigger jitter ?

fadc250.SB.doublepeak.pdf ( pdf)


  • FADC250 Signal: one channel - one event

fadc250.SB.onepeak.pdf ( pdf)