Characteristics
TDS400A
Theoretically achievable resolution with TDS400A Hi-res mode.
Signal
Acquisition System
Bandwidth - 200 MHz (TDS420A), 400 MHz
(TDS430A, TDS460A).
Channels - 4 (2 on TDS430A).
Sample
Rate - 100 MS/s on all channels.
Sensitivity - 1 mV to
10 V/div (with calibrated fine adjust).
Position Range - ±5
Divisions.
Offset Range - ±1 V from 1 mV/div to 99.5 mV/div;
±10 V from 100 mV/div to 995 mV/div; ±100 V from 1 V/div to 10 V/div.
DC
Gain Accuracy - ±1.5%.
Vertical Resolution - 8-Bit (256
levels over 10.24 vertical divisions).
Analog Bandwidth Selections
- 20 MHz, 100 MHz and full.
Input Coupling - AC, DC or
GND.
Input Impedance Selections - 1 MO in parallel with 15 pF
or 50 O (AC and DC coupling).
Maximum Input Voltage - 300 V
CAT II ±400 Vp. Derate at 20 dB/decade above 1 MHz. 1 megaohm or
GND coupled.
Channel Isolation - >100:1 at 100 MHz for any
two channels.
AC Coupled Low Frequency Limit - =10 Hz when
AC 1 MO coupled. =200 kHz when AC 50 O coupled.
Acquisition Modes
Peak
Detect - High frequency and random glitch capture. Captures glitches of
10 ns using acquisition hardware at all real-time sampling rates.
Sample
- Sample data only.
Envelope - Max/min values acquired
over one or more acquisitions, selectable from 2 to 2000, infinite.
Average
- Waveform averages selectable from 2 to 10,000.
Hi-res - Vertical
resolution improvement and noise reduction on low-frequency signals, e.g.
12-Bit at 10 ms/div and slower. Enhanced vertical resolution (>12-Bit) for
noise reduction, on low frequency signals. Make precise low-level signal measurements
(up to 5 µV) with differential amplifier (ADA400A).
Time Base System
Time
Bases - Main, Delayed.
Time/Division Range - 1 ns to 20 s/div.
Time
Base Accuracy - 0.015% over any interval =1 ms.
Record Length
(Real Time and Equivalent Time) - Sample points per channel: 500 to 120,000
points (with Opt. XL).
Pre-trigger Position - Selectable from
0 to 100% of record.
Triggering System
Triggers - Main,
Delayed.
Main Trigger Modes - Auto, Normal, Single Sequence.
Delayed
Trigger - Delayed by time or events.
Time Delay Range - 0 ns
to 20 s.
Events Delay Range - 2 to 10,000,000 events.
External
Rear Input - >1.5 kO; max input voltage is ±6 V (DC + AC peak).
Video
Trigger Types - NTSC, PAL, SECAM and custom; TV field, field 2 or both,
any line within a field. Line rates -10 kHz to 64 kHz, interlaced, noninterlaced,
composite.
Video Trigger Sensitivity - 0.6 divisions of composite
SYNC will achieve a stable display.
Display
Waveform Style
- Dots or vectors. Infinite and variable persistence from 250 ms to 10 s.
Gray
Scaling - With variable persistence selected, waveform points gradually
decay through 16 levels of intensity, providing "z-axis" information about
rapidly changing waveforms.
Update Rate - 200 ea. 500 point
wfms/s with infinite persistence mode selected.
Graticules - Full,
grid, cross hair, frame.
Format - YT and XY.
VGA
Out - Drives VGA display monitors.
Zoom
The zoom
feature allows waveforms to be expanded, compressed and positioned in both
vertical and horizontal axes. Allows precise comparison and study of fine
waveform detail without affecting ongoing acquisitions. When used with Hi-res
or Average acquisition modes, Zoom provides an effective vertical dynamic
range of 1000 divisions or 100 screens.
Measurement System
Automatic waveform measurements
|
|
Period
|
Frequency
|
High
|
Low
|
+ Width
|
- Width
|
Maximum
|
Minimum
|
Rise
|
Fall
|
Peak to Peak
|
Amplitude
|
+ Duty cycle
|
- Duty cycle
|
+ Overshoot
|
- Overshoot
|
Propagation delay
|
Burst width
|
Mean
|
Cycle mean
|
RMS
|
Cycle RMS
|
Area
|
Cycle area
|
Phase
|
-
|
Continuous update of up to four measurements on any combination
of waveforms.
Thresholds - Settable in percentage or voltage.
Gated
- Any region of the record may be isolated for measurement using vertical
bars.
Snapshot - Performs all measurements on any one waveform
showing results from one instant in time.
Cursor Measurements - Absolute,
Delta, Volts, Time, Frequency.
Cursor Types - Horizontal bars
(volts); vertical bars (time); paired; operated independently or in tracking
mode.
Waveform Processing
Waveform Functions - Interpolate-selectable
sin(x)/x or linear, Average, Envelope.
Advanced Waveform Functions
- FFT, Integration, Differentiation (Opt. XL).
Arithmetic Operators
- Add, Subtract, Multiply, Invert.
Autosetup - Single button,
automatic setup on selected input signal for vertical, horizontal and trigger
systems.
Waveform Limit Testing - Compares incoming waveform
to a reference waveform's upper and lower limits.
Computer Interface
GPIB
(IEEE 488.2) Programmability - Full talk/listen modes. Control of all
modes, settings and measurements.
Hardcopy/Desktop Publishing
Printer
- HP ThinkJet, Epson, PostScript, Interleaf, DeskJet, LaserJet, TIFF,
PCX, BMP (Microsoft Windows).
Plotter - HPGL.
Interface
- GPIB standard.
Hardcopy Interface - Centronics Type and
RS-232 (Opt. XL).
Storage
Waveforms - 120,000 waveform
points of nonvolatile storage (with Opt. XL).
Floppy Drive - 3.5 in.
1.44 MB or 720 KB MS DOS compatible (store waveforms, screen data, and setups)*1.
Setups
- 10 front-panel setups.
CRT
Type - 7 in. diagonal,
magnetic deflection. Horizontal raster-scan. P31 green phosphor.
Resolution
- 640 horizontal by 480 vertical displayed pixels.
Power Requirements
Line Voltage
Range - 90 to 250 VRMS.
Line Frequency - 48
to 440 Hz.
Power Consumption - 240 W max.
Environmental
and Safety
Temperature - Operating: 0 °C to +50 °C. Nonoperating:
-40 °C to +75 °C.
Humidity - Operating and nonoperating: Up
to 95% relative humidity at or below +40 °C; to 75% relative humidity from
+41 °C to +50 °C.
Altitude - Operating: 15,000 ft., nonoperating:
40,000 ft.
Electromagnetic Compatibility - 89/336/EEC.
Safety
- UL3111-1, CSA1010.1, EN61010-1, IEC61010-1.
Physical Characteristics
Dimensions
|
mm
|
in.
|
Weight
|
kg
|
lbs.
|
Height
|
165
|
6.4
|
w/acc. pouch
|
191
|
7.5
|
Width
|
362
|
14.25
|
Depth w/front cover installed
|
490
|
19.28
|
Depth w/handle extended
|
569
|
22.4
|
Net approximately
|
10.2
|
22.5
|
Shipping approximately
|
14.5
|
32
|
*1 Waveforms can be stored to file in MathCAD
and Spreadsheet (Excel®, Lotus 1-2-3®) formats for analysis.
Ordering
Information
TDS430A (order with Opt. XL)
Two-channel, 400 MHz
Digitizing Oscilloscope.
TDS420A (order with Opt. XL)
Four-channel,
200 MHz Digitizing Oscilloscope.
TDS460A (order with Opt. XL)
Four-channel,
400 MHz Digitizing Oscilloscope.
All Include: 1 probe per channel
(P6138A 10X Passive Probes); Video Trigger; Reference Manual (070-8035-03);
User Manual (070-8034-03); Performance Verification Document (070-8721-02);
power cord.
Please specify power plug when ordering.
Instrument
Options
Opt. XL - (Mandatory for all TDS400 models) 120,000
point record length; RS-232 and Centronics Hardcopy interfaces; Extended Waveform
Math; FFT; Integration, Differentiation; front cover and accessories pouch.
Opt.
1R - Rackmount Kit.
Probes
See Additional Accessories for
Viewing Hoods and Storage Compartments.
Differential Probes -
400 MHz,
30 dB CMRR. Order P6246.
High Voltage Probes -
2.5 kV,
25 MHz. Order P5100.
20 kV, 75 MHz. Order P6015A.
High Voltage
Differential Probes -
Up to 1300 V, 25 MHz. Order P5200.
Up
to 1300 V, 100 MHz. Order P5205.
Up to 5600 V, 50 MHz. Order P5210.
Passive
Probe 1x - Order P6101B.
Passive Probe 10x - Order P6138A.
FET
Probe - Order P6205.
SMT Probe - Order P6562A.
TTL
Logic Probe - Order P6408.
Optical Converters -
500 nm
to 950 nm. Order P6701B.
1100 nm to 1700 nm. Order P6703B.
DC/AC
Current Probe System - Order AM503S or TCP202 DC Coupled Current Probe.
Recommended
Accessories
Analog Differential Amplifier - 10 µV sensitivity.
Order ADA400A.
Current Measurement Capability - Order TDS400A
and AM503S, and appropriate probe.
Also see Additional Accessories for
Viewing Hoods and Storage Compartments.
Scope Cart - Order K420.
Rackmount
Kit - Order 016-1166-00.
Soft-sided Carrying Case - Order 016-1158-01.
Transit
Case - Order 016-1157-00.
Programmer's Manual - Order 070-9876-00.
Service
Manual - Order 070-9703-04.
Video Clamp - Order 013-0278-00.
Software
WSTRO
- WaveStar™ software for oscilloscopes, Windows 95/98/NT® application
for waveform capture, analysis, documentation and control from your PC.
WSTROU
- Upgrade from WSTR31 to WSTRO.
LVWIN95 - LabVIEW for Windows 95.
LWCVI95
- LabWindows/CVI for Windows 95.
Cables
GPIB -
1 meter.
Order 012-0991-01.
2 meters. Order 012-0991-00.
Power
Plug Options
Opt. A0 - US Plug, 115V, 60 Hz
Opt.
A1 - Euro Plug, 220V, 50 Hz
Opt. A2 - UK Plug, 240V, 50 Hz
Opt.
A3 - Australian Plug, 240V, 50 Hz
Opt. A4 - N. American
Plug, 240V, 50 Hz
Opt. A5 - Swiss Plug, 220V, 50 Hz
Opt.
A99 - No Power Cord
Opt. AC - China Plug, 50 Hz
International
power options required on instruments and selected accessories for operation
outside U.S. For operation outside U.S., specify A1-A5 power options. See
General Customer Information pages for additional description.
Service
Opt.
C3 - Calibration Service 3 Years
Opt. C5 - Calibration
Service 5 Years
Opt. D1 - Calibration Data Report
Opt.
D3 - Calibration Data Report 3 Years (with Option C3)
Opt.
D5 - Calibration Data Report 5 Years (with Option C5)
Opt.
R5 - Repair Service 5 Years
See General Customer Information for
further information.